Nellcor Puritan Bennett 840 PB840 Ventilator Medtronic Mechanical Ventilator PB 840 – Faulty

$1,700.00 inc. GST

Out of stock

SKU: 20032711-3714RM Category:

Description

Used: Tested, Faulty

Photos are of the actual item

AS IS, No Warranty

Accessories – Air & O2 Hoses Only

Weight (kg) – 80

Size (L x W x H) (cm) – 75 x 75 x 145

Additional Comments

This ventilator was purchased from a local government agency.

This ventilator is NOT in working order.

Upon testing the ventilator fails self test due to the below errors:

* FS0001 – O2 Flow Sensor Cross Check Failed (Q1)

* FS0003 – Air Flow Sensor Cross Check Failed (Q2)

* EST Passed.

* SST Failed: ‘Flow Sensor Calibration Failed’.

This could be a simple fix by replacing the Q1 & Q2 sensors [from USD $67.50 each].

Ventilator hours : 29,967.

Options: BiLevel, PAV+

Additional photos are available upon request.

Manufacturer’s Overview

https://www.medtronic.com/covidien/en-us/products/mechanical-ventilation/puritan-bennett-840-ventilator.html

FREIGHT QUOTE:

Yes, we ship WORLDWIDE!

We can ship by Air or Sea.

Please be aware that freight for this may be expensive – Please contact us with the following information for a detailed freight quote:

* Country, Suburb/town & Postcode/zip code?

* Is this a residential or commercial/business address?

* If commercial/business, is there a forklift to unload?

* Can you collect it from a local freight depot? (This is usually the cheapest option)

20032711-3714RM

AIOSM

Additional information

Weight110 kg
Dimensions80 × 80 × 165 cm
Shelf1-04-A02A