Accessories – Air & O2 Hoses Only
This ventilator was purchased from a local government agency.
This ventilator is NOT in working order.
Upon testing the ventilator fails self test due to the below errors:
* FS0001 – O2 Flow Sensor Cross Check Failed (Q1)
* FS0003 – Air Flow Sensor Cross Check Failed (Q2)
* EST Passed.
* SST Failed: ‘Flow Sensor Calibration Failed’.
This could be a simple fix by replacing the Q1 & Q2 sensors [from USD $67.50 each.
This ventilator is also missing the GUI to BDU cable – REF 4-075864-00 – See last 2 images [from USD $40.00
* Utilised a cable from another unit to test
Ventilator hours : 20,535.